Cov Txheej Txheem Ntau Lawm & Kev Tswj Xyuas Zoo Ntawm PTC Koob
Jul 06, 2026
https://admin1.seo.com.cn/CustomerAdmin/S_News/Create?ru=%2FCustomerAdmin%2FS_News%2F
Txawm hais tias nws qhov yooj yim yooj yim, Chiba koob raug tsim ua ke nrog kev ua cov raj ua tiav, micro- sib tsoo, kev kho saum npoo, thiab cov cuab yeej kho mob tsis muaj zog QC.
① Raw Khoom Xaiv
Ncej:Seamless capillary raj rau ASTM F138 / F139 (316L lossis 304 qis-carbon stainless); puab nto feem ntau electropolished hauv raw khoom
MRI- tau tshaj versions:Nitinol (NiTi) los yog Titanium raj; Ni ion tso tawm yuav tsum ua raws li ISO 10993-1
Hub: Kev kho mob- qib polycarbonate (PC) lossis ABS - pob tshab, EO / gamma kam rau ua, DEHP-dawb
② Tube Txiav & Ncaj
Siab -precision feed cutter → ntev nrog kev ua haujlwm pub dawb; txiav lub ntsej muag perpendicularity yuam kev<0.05 mm
Peb - yob straightening → puv -length deviation Tsawg dua los yog sib npaug rau 0.3 mm los tiv thaiv intra- txheej txheem deflection
③ Lumen Conditioning & Deburring
Sab hauv txoj kab uas hla ua tiav los ntawm ntsaws- kos duab / hlawv
Ob qhov kawg chamfered kom tshem tawm burrs → tiv thaiv stylet jamming & cov ntaub so ntswg tearing
④ Koob Tip Sib Tsoo(Cov txheej txheem tseem ceeb)
Sab nraud grind rau classic25℃bevel(los yog ib spec); Secondary nyem grind (rov qab-txiav) xaiv los txo qhov kev quab yuam
Fine pob zeb diamond abrasive / paste polish → microscope- tshuaj xyuas rau symmetric bevel, tsis muaj dov ntug, tsis muaj micro-chipping
100% kev tshuaj xyuas qhov muag lossis AOI sampling ib SPC
⑤ Ncej Marking & Surface Treatment
Laser-etched centimeter kawm tiav los ntawm lub ntsis; number case los yog laser xim sau kom nyeem tau
Yeem: passivation (stainless) lossis npub-coat (Teflon® nyias zaj duab xis) kom txo qis
Echo marker: micro-indentation lossis localized laser kev ntxhib los mos ze ntawm lub taub; platinum / titanium nplhaib ntawm hwm kab
Nitinol shafts: electropolish + acid passivation kom tshem tawm nto - qhib nickel txheej
⑥ Hub Assembly & Bonding
Koob raj crimped/adhered rau hauv Luer-Xauv los yog Luer-Slip hub nrog kho mob epoxy los yog UV nplaum; torque-tested (feem ntau ntau dua los yog sib npaug rau 0.5 N·m tsis yog-loosing)
Stylet (yog tias suav nrog) sib sau ua ke - stylet tip protrudes 0.5-1 mm dhau ntawm cannula taub los pab cov ntaub so ntswg nkag & tiv thaiv lumen clogging
Kev ntsuas dej: huab cua siab siv → submerged → tsis muaj npuas emission
⑦ Kev Tu & Kev Tswj Pwm
Multi-theem purified dej yaug, ntxuav ultrasonic kom tshem tawm cov sib tsoo sib xyaw & hlau nplua
Thaum kawg yaug kuaj ibUSP<788> particle teeb meem txwv → mitigates micro-embolism txaus ntshai
⑧ Sterilization & Ntim
Feem ntauEthylene Oxide (EO)ua kom tsis muaj menyuam; validate koob tshuaj, aeration voj voog, EO / ECH residuals (EO Tsawg dua lossis sib npaug ntawm 4 µg / khoom siv, ECH Tsawg dua lossis sib npaug rau 9 µg / khoom siv rau ISO 10993-7)
Sterility Assurance Level (SAL)=10⁻⁶
Tyvek® hau/blister los yog tev-pob hnab; sab nrauv thawv nrog LOT, tas sij hawm, Gauge × Length, sau npe No.
⑨ Kev Tswj Xyuas Kev Tswj Xyuas Zoo & Kev Xeem Tso Tawm
puvISO 13485 Kev Tswj Xyuas Kev Lag LuamQMS; Cov txheej txheem tshwj xeeb (sib tsoo, sib dhos, ua kom tsis muaj menyuam) validated ntawm IQ / OQ / PQ
Cov khoom tiav sampling:
Dimensional (OD/ID/bevel kaum/ntev - pom kev ntsuas)
Lub tswv yim morphology (microscopy)
Hub rub- tawm lub zog
Lumen patency / ntws
Sterility, pyrogen (LAL lossis luav)
Biocompatibility (ISO 10993 cytotoxicity / sensitization / hemolysis - qauv ntau)
Kev xa tawm: FDA 21 CFR 820 (US) / EU MDR (CE) raws li muaj
OEM / ODM customization (2D / 3D kos duab lossis qauv) feem ntau suav nrog Kev Ntsuas, qhov ntev, bevel kaum sab xis, hub xim code, echo tip muaj, thiab suav nrog micro-cov lus qhia teeb. Sampling 4-8 lub lis piam; lub sij hawm ua cov hmoov txhuas sib txawv nrog ntim. Cov ntsiab lus tseem ceeb ntawm tus neeg muag khoom tshawb xyuas: siv tau ISO 13485, NMPA (lossis sib npaug) sau npe suav nrog kev npaj siv, batch traceability, thiab keeb kwm tsis zoo tshwm sim.